Since 1992 전통의 환경측정 장치 전문기업
홈 > 수입공급제품
수입공급제품
Arizona instrument
|
|
|
|
|
[Jerome 651] |
H2S 분석장치
AMETEK Arizona Instrument
https://www.azic.com/ |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detection Range: | 3 ppb (0.003 ppm) to 50 ppm | Resolution: | 0.001, 0.01, 0.1, and 1 ppm | Accuracy: | ±0.005 ppm at 0.05 ppm ±0.05 ppm at 0.5 ppm ±0.5 ppm at 5 ppm ±2 ppm at 25 ppm | Sample Intervals: | 1, 2, 5, 10, 15, 30, 60 or 120 minutes | Regeneration Intervals: | 6, 12, 24 and 48 hours | Flow Rate: | 150 mL/min | Operating Environment: | -40 to +55°C, non-explosive, 0-100% relative humidity | Result Units: | ppm, ppb | Data Storage Capacity: | 50,000 readings | Power Requirements (651 Unit): | 100-120 VAC, 50/60 Hz, 7 A or 210-240 VAC, 50/60 Hz, 4 A | Control Board Fuse: | Fast-acting IEC 1.6 A 250 V, 5 mm x 20 mm | Case Construction: | Polyester with insulation to reduce heat transfer | Battery (Detatched 631-X Unit): | Rechargeable NiCad 6 hour life | Estimated Sensor Life: | 3 to 6 months depending on operating conditions | Display: | 8 digit alphanumeric LCD | Output: | Spectra Radio Analog: 4-20 mA passive current loop | Dimensions: | 26″ W x 25″ H x 9″ D (67 cm W x 64 cm H x 23 cm D) | Weight: | 52 lbs. (24 kg) | Warranty: | 1 year, factory parts and labor except for 631-X sensor |
Spectra Radio Modem SpecificationsFrequency: | 2.4 – 2.4835 GHz (2400 MHz) 49 selectable channels | Serial Interface: | RS-232 | Power Requirements: | 10-30 VDC, 450 mA | Operating Temperature: | -40°C to 75°C | Range: | Up to 4 miles (6.4 km) “line of sight.” Longer distances may be obtainable through use of a high gain antenna or by using the optional Radio Repeater |
|
|
|
|